Resumen
The design of a low-power-integrated potentiostat that measures electric current intended for a Chagas disease detection application is presented. This circuit can generate and measure currents from 1 to 10 μA with a relative error under 5%. The remaining part of the circuit consumes only 1.5 μA. The circuit was designed using the 0.6 μm XC06 technology from the X-FAB, with a total silicon area of 0.24 mm2 (without pads).
Idioma original | Inglés |
---|---|
Páginas (desde-hasta) | 2299-2313 |
Número de páginas | 15 |
Publicación | International Journal of Circuit Theory and Applications |
Volumen | 46 |
N.º | 12 |
DOI | |
Estado | Publicada - dic. 2018 |